[. . . ] REPLACEMENT POWER SUPPLY KIT for RPM3 or molbox RFM Installation Instructions REPLACEMENT POWER P/N 401903, INCLUDES QTY 1 1 1 2 PART NO 102705 102704 102722-Z 102721-Z DESCRIPTION SUPPLY KIT, Power Supply, +5V, +12V Plastic Standoff, Locking Screw, Nylon, M3 x 12 Washer, Nylon, M3 This is a replacement power supply kit for RPM3 or molbox RFM. [. . . ] Locate the tabs of the plastic standoff and squeeze them together to release the standoff from the main board. It is not necessary to remove the plastic standoff from the old power supply, a new standoff is included in this kit Remove and discard the nylon washers located between the metal mounting tabs and the power supply. Figure 1 Top View 1. Front of Unit INSTALLING THE NEW POWER SUPPLY: Install the new plastic standoff into the new power supply. Position the power supply in the correct mounting location with the plastic standoff directly above the mounting hole on the main board. Press down on the power supply to captivate the plastic standoff to the main board. Install a nylon washer between the power supply and the metal mounting tab, two (2) places, (see Figure 3). [. . . ] Document 560047 020724 DH Instruments, Inc. 4765 East Beautiful Lane Phoenix AZ 85044-5318 USA Tel 602. 431. 9100 Fax 602. 431. 9559 dhi@dhinstruments. com www. dhinstruments. com [. . . ]